Washer-disinfectors. Part 7: Requirements and tests for washer-disinfectors employing chemical disinfection for non-invasive, non-critical thermolabile medical devices and healthcare equipment (ISO 15883-7:2016)
Tuote ladattavissa hetiToimitusaika on noin 1 - 2 työpäivääToimitusaika on 3 - 5 työpäivää.
ISO 15883-7:2016 specifies the particular requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of reusable items such as the following: a) bedframes; b) bedside tables; c) transport carts; d) containers; e) surgical tables; f) sterilization containers; g) surgical clogs; h) wheelchairs, aids for the disabled. ISO 15883-7:2016 also specifies the performance requirements for the cleaning and disinfection of the washer-disinfector and its components and accessories which may be necessary in order to achieve the required performance. Devices identified within the scopes of ISO 15883-2, ISO 15883-3, ISO I5883-4, and ISO 15883-6 do not fall within the scope of this part of ISO 15883. In addition, the methods are specified, as well as instrumentation and instructions required for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control, and monitoring, as well as requalifications required to be carried out periodically and after essential repairs. NOTE WDs corresponding to this part of ISO 15883 can also be used for cleaning and chemical disinfection of other thermolabile and reusable devices as recommended by the device manufacturer. The performance requirements specified in this part of ISO 15883 may not ensure the inactivation or removal of the causative agent(s) (prion proteins) of Transmissible Spongiform Encephalopathies.
Tämän julkaisun valmistelusta Suomessa vastaa Palvelualojen työnantajat PALTA ry, puh. 040 579 4393.
Sisällysluettelo
Foreword
Introduction
1 Scope
2 Normative references
3 Terms and definitions
4 Performance requirements
4.1 General
4.2 Cleaning
4.3 Disinfection
4.4 Final rinsing
4.5 Self-disinfection
4.6 Drying
4.7 Water treatment equipment
5 Mechanical requirements
5.1 Materials — Design, manufacture, and assembly
5.2 Process verification
6 Testing for conformity
6.1 General
6.2 Test load
6.3 Water used for rinsing following disinfection
6.4 Load dryness
6.5 Thermometric tests
6.6 Chemical dosing tests
6.7 Tests of cleaning efficacy
6.8 Test of disinfection efficacy
7 Documentation
8 Information to be provided by the manufacturer
9 Marking, labelling, and packaging
10 Information to be requested from the purchaser by the manufacturer
Annex A Summary of test programmes (normative)
Annex B Methods for microbiological evaluation of disinfection of liquid transport system (normative)
Annex C Tests for microbiological contamination of post-disinfection rinse water (normative)
Annex D Preparation and evaluation of indicators for microbiological testing of the efficacy of chemical disinfection of the load (normative)
Annex E Examples of test locations for the tests with biological indicators (informative)
EN 10088-1:2014 Stainless steels - Part 1: List of stainless steels
EN 10088-1:2023 Stainless steels - Part 1: List of stainless steels
EN 10088-2:2014 Stainless steels - Part 2: Technical delivery conditions for sheet/plate and strip of corrosion resisting steels for general purposes
EN 10088-1:2014 Stainless steels - Part 1: List of stainless steels
EN 10088-1:2023 Stainless steels - Part 1: List of stainless steels
EN 10088-2:2014 Stainless steels - Part 2: Technical delivery conditions for sheet/plate and strip of corrosion resisting steels for general purposes
EN 10088-2:2024 Stainless steels - Part 2: Technical delivery conditions for sheet/plate and strip of corrosion resistant steels for general purposes
EN 61010-2-040:2005 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-040: Particular requirements for sterilizers and washer-disinfectors used to treat medical materials
EN ISO 15883-6:2015 Washer-disinfectors - Part 6: Requirements and tests for washer-disinfectors employing thermal disinfection for non-invasive, non-critical medical devices and healthcare equipment (ISO 15883-6:2011)
IEC 61010-2-040:2005
ISO 15883-1:2006 Washer-disinfectors -- Part 1: General requirements, terms and definitions and tests
ISO 15883-1:2006/Amd 1:2014
ISO 15883-4:2008 Washer-disinfectors -- Part 4: Requirements and tests for washer-disinfectors employing chemical disinfection for thermolabile endoscopes
ISO/TS 15883-5:2005 Washer-disinfectors -- Part 5: Test soils and methods for demonstrating cleaning efficacy